GE Healthcare – Lunar DPX IQFull Body Bone Densitometer, The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better. service manual for lunar dpx. Sat, 20 Oct GMT service manual for lunar dpx pdf – DPX Series. Service repair lunar dpx iq service manual – free. Frank’s Hospital Workshop. – service manual lunar dpx nt LUNAR DPX Operators . Manual, lunar dpx iq service manual pdf Gluer.

Author: Moogujar Fenrile
Country: Nicaragua
Language: English (Spanish)
Genre: Relationship
Published (Last): 9 February 2015
Pages: 191
PDF File Size: 7.88 Mb
ePub File Size: 19.25 Mb
ISBN: 351-5-25554-549-5
Downloads: 51189
Price: Free* [*Free Regsitration Required]
Uploader: Nadal

We have it in stock if you need one.

Display in Community Feed. Advertise Increase Your Brand Awareness.

Service manual for GE Lunar DPX-NT densitometer

Just remember that the detector is not properly responsive during QC and exams as well. I have a problem with this team is that everything works, from linar source to the Oink card, but does not open the solenoid of the collimator.

You likely have a bad detector. Tengo un problema con este equipo, resulta que todo funciona, desde la fuente, hasta la tarjeta Oink, pero no abre el solenoid del colimador.

If you would like a quote feel free to email me at ben servicw.

Had the same problem was bad detector. There’s some thing i can do, and can someone send me the service manual, also?

  ANDRES LOPEZ EL CARTEL DELOS SAPOS PDF

Add to My Bench. You can add Community Subscriptions in the search bar that says “Subscribe to more communities You can sometimes get past luhar by using the x-ray test mode to saturate the detector. This is a very common failure on any of the DPX models. Sign In or Register. Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment.

Ad Statistics Times Displayed: Remove Subscription Confirm Remove. The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better image quality.

Also, as dpxx mentioned, this is a part you do not need any specialty tools to replace if the alignment has not changed.

GE Healthcare – Lunar DPX IQ Community, Manuals and Specifications | MedWrench

Display in Community Feed. Greetings and thanks for your servicd Reply. If active manually if the solenoid is activated, similarly, if I activate the switch, achievement cheat the system and I do not make the mistake, but if only the active from the system, it does nothing. The language is no problem, actually is perfect. On the momento the bmd is in 1, L2-L4,normal ranging 1, to 1,the last reading before of the fail was 1, You may be getting improper results on the patient.

The detector is easy to replace but you will likely have to run the peak alignment. Please review our Privacy Policy for more details.

  ISO 7823-1 PDF

Having worked for the Lunar business for many years we have seen these failures numerous times. Hi Adriano, As mentioned, if nothing has changed recently with alignment, then your detector has deteriorated.

GE Healthcare – Lunar DPX IQ Community, Manuals and Specifications | MedWrench

Tengo un problema con este equipo, resulta que todo funciona, desde la fuente, hasta la tarjeta Oink, pero no abre el solenoid del colimador.

This site uses cookies. By continuing to browse the site you are agreeing to our lunaar of cookies. The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better image quality. Additionally, we season the detectors in a system prior to shipping to ensure quality and to save you from having to complete that process.

Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment. My Bench Order History Sign out. My Bench Order History Sign out.